C 7685.001 E.Conductivity resistivity analyzer
Specifications
- Conductivity range: see table on the data sheet
- Temperature range: -10/+110 °C, accuracy 0.1°C
- Analog output: 0-20 or 4-20 mA isolated. Programmable within the selected scale
- 2 set-point with min/max function, hysteresys and delay time programmable
- Min/max and set points timing alarm relay
- Dual software filter for more steady measures
- 3 software access levels, keyboard lock and access code number
- Automatic recognition of the standard solutions
- Operating mode: automatic and manual
- Temperature readout
- Ultrapure water temperature coefficient table stored in the microcomputer
- Operating mode: automatic and manual
- Continuous or flashing alarm relay
- Power supply: 110/220 V, 5 VA
- Dimensions: 96x96x155 mm. DIN 43700
Technical Specifications
in addition to those common in the series 7685
Operating mode
Automatic/Manual
E.C. Cells
* K = 0.01/0.10/1.0/10 cm-1
* Input scales: 200.0 nS/20.00 mS (see table) readout in Mohm
* Autoranging: ON/OFF
Zero adjustment: +/- 10 %
Sens. adjustment: 60/160 %
* Temp. reference: 10/25 °C
* ATC coefficient: 0/3.50 %/°C
* Software filter 90%RT: 0.4/20.0 s for small/large variations
Display resolution: 1/1000 at 20 °C
Temperature
Input: RTD Pt100 2/3 wires connection
Measuring and compensation range: 0/+100 °C
Resolution: +/- 0.1 °C
Zero adjustment: +/- 2 °C
ManualTemp. comp: 0/100 °C
Options
091.3713 – dual analog programmable and isolated output.
The operator may select an output for Temperature.
091.4143 9 – /36VDC power supply